Operation Verification

ESG Family Signal Generators

Verification Procedures

2-6

Calibration Guide

Configure the Signal Generator

1. Press

Frequency

. Enter 277 using the numeric keypad and press the

kHz

terminator

softkey.

2. Press

Amplitude

. Enter

15 using the numeric keypad and press the

dBm

terminator

softkey.

Configure the Spectrum Analyzer

1. Preset the spectrum analyzer.

2. Set the analyzer to external 10 MHz reference.

3. Set the center frequency to 277 kHz.

4. Set the frequency span to 100 Hz. (This will result in a 1 Hz resolution bandwidth with

a digital filter.)

Measure the Power Level Accuracy

1. Connect the signal generator’s RF OUTPUT through the 6 dB attenuator to the

spectrum analyzer’s RF input.

2. Transfer the power level results for each frequency at

15 dBm from the Measured

Power column in

Table 2-2

or

Table 2-3

as appropriate, to the Power Meter Reading for

15 dBm column in

Table 2-4, “Power Level Accuracy Setup 2 (All Signal Generators),”

on page 2-15.

3. On the spectrum analyzer, select

Marker Normal Mode

and then select the

Peak Search

function. This activates the marker and sets it to the signal peak.

4. On the spectrum analyzer, ensure that the marker is at the signal peak and use the

MKR->

menu to set the marker to the reference level. If necessary, select

Peak Search

to

ensure that the marker is at the signal peak.

5. With the marker at signal peak, select the

Marker Delta

function. This will set the marker

to measure relative amplitude from a reference of 0 dB. If the marker does not read
0 dB, press

Marker Normal

>

Peak Search

>

Marker Delta

until the marker reads 0 dB.

6. Decrease the signal generator amplitude in 10 dB steps as shown in

Table 2-4

. At each

10 dB step, select

Peak Search

to ensure that the marker is at the signal peak.

7. Measure the power levels shown in the Power Level Setting (dBm) column of

Table 2-4

for the current frequency, and record the values in the Spectrum Analyzer Marker (dB)
column.

8. Return the spectrum analyzer reference level and the signal generator amplitude to

15 dBm. Set the signal generator frequency and the spectrum analyzer’s center

frequency to the next frequency listed in

Table 2-4

and repeat the process from step 3.

Continue until all of the frequencies have been measured and recorded (to the
maximum frequency of your signal generator).