Fm/pm out-of-band calibration, Description, Required test equipment – Atec Agilent-8563E User Manual

Page 119: Procedure

background image

Adjustments

ESG Family Signal Generators

FM/PM Out-of-Band Calibration (ESG-AP and ESG-DP only)

5-22

Calibration Guide

FM/PM Out-of-Band Calibration (ESG-AP and ESG-DP
only)

Description

This adjustment calibrates the FM/PM out-of-band paths on the Fractional-N and YO
driver modules. The internal ABUS is connected to the phase lock loop integrator node on
each board, and the attenuators and DAC are adjusted to minimize the voltage. This
ensures that the in-band and out-of-band paths are matched. The

“FM In-Band DAC

Offset Calibration”

on page 5-15 must be performed prior to this adjustment.

Required Test Equipment

• HP/Agilent 3458A Digital Multimeter

• HP/Agilent 33120A 15 MHz Function/Arbitrary Waveform Generator

Procedure

Figure 5-18 FM/PM Out-of-Band Calibration Setup

1. Connect the equipment as shown above.

2. Preset all of the equipment.

3. Follow the instructions as they appear on the controller’s display.