Alc modulation driver bias calibration, Description, Required test equipment – Atec Agilent-8563E User Manual

Page 130: Procedure

background image

ESG Family Signal Generators

Adjustments

ALC Modulation Driver Bias Calibration

Calibration Guide

5-33

ALC Modulation Driver Bias Calibration

Description

This test adjusts the ALC MOD DRV BIAS DAC on the output board. This DAC is used to
control the bias current to the ALC modulator driver. It is primarily used to accommodate
unmatched V

BE

values in the drives, but has a strong influence on modulator gain. This

test adjusts the DAC until the ALC modulator gain is balanced around its nominal design
center. The results are stored as a calibration array in the signal generator’s firmware.

Required Test Equipment

• None

Procedure

1. Preset the signal generator.

2. Follow the instructions as they appear on the controller’s display.