Atec Agilent-8563E User Manual
Page 44
ESG Family Signal Generators
Service Software
Running the Service Support Software
Calibration Guide
3-23
3. Once the test is highlighted, identify all test equipment listed in the Required Test
Equipment box. To identify test equipment:
a. Select the type of device from the Device Type list.
b. Select the model number from the Model list for the device.
c. Select the device’s serial number from the Available Test Equipment list.
d. Select the Add button to add the device to the list in the Selected Test Equipment
box.
The following buttons are available for the Selected Test Equipment box:
Add
Copies the tests and adjustments highlighted in the Available Test
Equipment box to the Selected Test Equipment box.
Remove
Removes the highlighted test equipment from the Selected Test
Equipment box.
Clear
Removes all of the test equipment from the Selected Test Equipment
box.
NOTE
If necessary, the test equipment GPIB address can be changed after it is
added to the Selected Test Equipment box. Change the GPIB address by
pressing the right arrow on the keyboard until the GPIB address selection in
the Selected Test Equipment box is selected. (The GPIB address is selected
when it has a dark box around the selection.) Then, type the new GPIB
address and press Enter to change the address.
e. Repeat the previous steps for each test equipment instrument.
4. Once the test equipment has been identified, select the Add button located above the
Selected Tests box to add each of the highlighted performance tests or adjustments. The
tests or adjustments that were highlighted are now displayed in the list in the Selected
Tests box.
The following buttons are available for the Selected Tests box:
Add
Copies the tests and adjustments highlighted in the list of tests and
adjustments to the Selected Tests box.
Remove
Removes the highlighted tests and adjustments from the Selected Tests
box.
Clear
Removes all of the tests and adjustments from the Selected Tests box.