Adjustments

ESG Family Signal Generators

AM Gain Calibration

5-40

Calibration Guide

AM Gain Calibration

Description

This test adjusts the gain of the AM circuitry to provide a 10 dB drop for a 1 V input signal.
First a power level is set and a one-volt signal is connected to the input. AM is enabled and
the AM DACs are adjusted for an exact 10 dB drop. The resulting DAC value is stored as a
calibration array.

Required Test Equipment

• HP/Agilent 438A Power Meter

• HP/Agilent 8482A Power Sensor

• HP/Agilent 33120A 15 MHz Function/Arbitrary Waveform Generator

• HP/Agilent 3458A Digital Multimeter

Procedure

Figure 5-34

AM Gain Calibration Setup

1. Connect the equipment as shown above.

2. Preset all of the equipment.

3. Follow the instructions as they appear on the controller’s display.