beautypg.com

Measuring residual fm, Figure 4-37 – Atec Agilent-8563E User Manual

Page 90

background image

ESG Family Signal Generators

Performance Tests

21. Measuring Phase Noise and Residual FM (ESG-AP and ESG-DP Series Signal Generators)

Calibration Guide

4-41

Figure 4-37

Measuring Residual FM

Residual FM is closely related to phase noise. Good phase noise typically implies good
residual FM. Directly measuring residual FM at very low levels is difficult, but the
HP/Agilent E5500 system can integrate the phase noise results to determine an accurate
value. This is done under the E5500 software Trace Integration menu by selecting a data
type of

Snu(f)

or

Spectral density of frequency fluctuations

.

In the ESG-P series, residual FM is defined only within a 300-3000 Hz bandwidth and
with a specified frequency response within this band. That frequency response is specified
by a CCITT filter characteristic. In the E5500 system, software revision 5.0 or higher, use
of this filter response can be achieved by simply checking the

CCITT weighting

box under

the

Trace Integration

menu.

If you use the older HP/Agilent 3048 phase noise system, there is no direct way to measure
residual FM, but because residual FM and phase noise are so closely related and
interdependent, it is safe to assume that if the DUT’s phase noise meets specification,
looks typical (between 300 and 3000 Hz), and shows no unexpected large spikes in the
response, then it will most likely also meet the residual FM specification.