Adjustments

ESG Family Signal Generators

Burst Audio Path Gain Calibration

5-28

Calibration Guide

Burst Audio Path Gain Calibration

Description

This test adjusts the scaling of the linear burst audio path until a -0.99 Vdc input to the
EXT1 connector results in a 40 dB drop in power relative to 0.00 Vdc input when linear
burst is activated. The AM 1 DAC value is changed until this result is achieved and its
value is stored as a calibration array.

Required Test Equipment

• HP/Agilent 3458A Digital Multimeter

• HP/Agilent 33120A 15 MHz Function/Arbitrary Waveform Generator

• HP/Agilent 8563E Spectrum Analyzer

Procedure

Figure 5-24

Burst Audio Path Gain Calibration Setup

1. Connect the equipment as shown above.

2. Preset all of the equipment.

3. Follow the instructions as they appear on the controller’s display.