ESG Family Signal Generators

Adjustments

FM/PM YO Frequency Compensation Calibration (ESG-AP and ESG-DP only)

Calibration Guide

5-23

FM/PM YO Frequency Compensation Calibration
(ESG-AP and ESG-DP only)

Description

This adjustment calibrates the FM/PM YIG oscillator frequency compensation latches on
the YO driver assembly. These latches adjust the flatness of the YO FM/PM paths for rates
greater than 100 kHz. The

“FM In-Band DAC Offset Calibration”

on page 5-15 and the

“FM/PM Out-of-Band Calibration (ESG-AP and ESG-DP only)”

on page 5-22 must be

performed prior to this adjustment.

Required Test Equipment

• HP/Agilent 8563E Spectrum Analyzer

• HP/Agilent 33120A 15 MHz Function/Arbitrary Waveform Generator

Procedure

Figure 5-19 FM/PM YO Freq. Compensation Setup

1. Connect the equipment as shown above.

2. Preset all of the equipment.

3. Follow the instructions as they appear on the controller’s display.