Performance Tests

ESG Family Signal Generators

6. Phase Modulation Frequency Response

4-16

Calibration Guide

6. Phase Modulation Frequency Response

This automated test verifies the phase modulation frequency response specifications. The
equipment measures the variations in phase deviations due to changes in the applied

Φ

M rate; dc to 100 kHz. The response is relative to a 1 kHz reference at the same phase

deviation, and expressed in dB. Each frequency is tested with the following sequence:

1. A reference signal is measured.

2. The set deviation is measured for each applied rate change.

3. Each deviation measurement is compared to the reference signal deviation.

The comparison is a calculation of the difference in the deviations relative to the reference
rate. The relative value is the

Φ

M frequency response.

Recommended Equipment

• HP/Agilent 8902A Measuring Receiver Option 010

• HP/Agilent 8904A Function Generator

Equipment Setup

Figure 4-14 Phase Modulation Frequency Response Test Setup