Performance Tests

ESG Family Signal Generators

Support Software

4-6

Calibration Guide

10.To save the test results to a file, make the following selections (

Figure 4-4

):

a. Click on the Drives window (item 20) and select the drive where you want to save the

test results.

b. Click on the Save file as type window (item 21) and select the file type for saving the

test results.

c. Click on the File name window (item 22) and type in a name for the test results file.

11.Click OK (item 23).

Figure 4-4

The Save As Window

12.In the Main Test and Results window (

Figure 4-5

), verify that the selected DUT (item

24), serial number (item 24), and selected tests (item 25) are correct. Then click Run.

The selected performance tests are listed in the order in which they will run. You
cannot change this order.

During the testing sequence the window shows the following information:

• the model and serial number of the signal generator under test (item 24)

• the list of selected performance tests with the current test highlighted (item 25)

• the performance test currently running (item 26)

• the data points taken (item 27)

• the total number of data points that will be measured during the test (item 28)

• the number of points that have currently passed the test (item 29)

• the number of points that have currently failed the test (item 30)

• the lower limits (LL) (item 31) and upper limits (UL) (item 32) for each test point

• the measured results (item 33) and the pass/fail (P/F) (item 34) indication