Performance Tests

ESG Family Signal Generators

11. Timebase Aging Rate (ESG-AP, ESG-DP or Option 1E5 only)

4-26

Calibration Guide

4. If the signal drifts a full cycle (360

°

) in less than 2 minutes, refer to

Chapter 5,

“Adjustments,”

and perform the “Internal Reference Oscillator Adjustment.” After the

adjustment, restart this performance test.

5. Watch the oscilloscope display and monitor the time. Notice the time required for a 360

°

phase change and record this time as T1.

6. Wait 3 to 24 hours. Record the time that you waited as T2.

7. Repeat steps 1 through 6. Notice the time required for a 360

°

phase change and record

this time as T3.

8. Calculate the aging rate as follows:

Aging Rate = (1 cycle/10 MHz) (1/T1

1/T3) (24 hours/T2)

Example:

T1 = 351 seconds

T2 = 3 hours

T3 = 349 seconds

= (1 cycle/10 MHz) (1/351s

1/349s) (24h/3h)(1.306

×

10

-11

per day)

9. Write the test results on the performance test record located at the end of this chapter.

Compare the results to the limits in the test record.

NOTE

If the absolute frequency of the standard and the timebase oscillator are
extremely close, you can reduce the measurement time (T1 and T3) by
measuring the time required for a phase change of less than 360

°

. In step 4,

change 1 cycle to 0.5 cycle for 180

°

, or 0.25 cycle for 90

°

.