To set the Over-Current Protection:

To set the Under-Current Protection:

Step # Do This

You Will See

1

OVP SET 0V

2

Press repeatedly until the

desired setting appears.

OCP SET 0A

3

OCP SET 0A

4

Use the Current knob, the numeric
keypad, or the arrow keys to enter a
value (0–103% of the unit’s rated
current).

OCP SET ####A

5

S/D if trip? N

6

Use the arrow keys to select Yes or No.
For this example, Yes is selected.

S/D if trip? Y

7

This setting is saved and the display
returns to its default operating mode.

PROT SET

ALARMS

PROT SET

ALARMS

ENTER

ENTER

ENTER

Step # Do This

You Will See

1

OVP SET 0V

2

Press repeatedly until the desired setting

appears.

UCP SET 0A

3

UCP SET 0A

4

Use the Current knob, the numeric keypad, or the
arrow keys to enter a value. The value must be
between 0–103% of the unit’s rated current.

UCP SET ####A

5

S/D if trip? N

6

Use the arrow keys to select Yes or No. For this
example, Yes is selected.

S/D if trip? Y

7

This setting is saved and
the display returns to its
default operating mode.

PROT SET

ALARMS

PROT SET

ALARMS

ENTER

ENTER

ENTER