Exit calibration mode

TM-XDOP-01XN

B–15

Exit calibration mode

When you have completed calibration, hit the EXIT key.

The SCPI command is:

CAL:STAT OFF, "0000"

Restore Factory Calibration

To restore the unit to the calibration constants set at the factory:

The SCPI command is:

CALibration:RESTore

Step # Do This

You Will See

1

Output V Cal

2

or

Scroll to select Factory Cal.

Factory Cal

3

Select Y to Restore the factory
calibration. Selecting N will exit.

Restore? Y

Important:

This procedure should not be used in place of regular calibration, but may

be useful to restore the unit to an operational state in case of failure.