Scpi – Xantrex Technology XDC 20-600 User Manual
Page 212

Calibration (6000 Watt only)
B–4
TM-XDOP-01XN
•
ANLG V PGM 10V: Calibrate 10V analog programming interface for setting
voltage output.
•
ANLG V RB 10V: Calibrate 10V analog programming interface for
monitoring voltage output.
•
ANLG I PGM 10V: Calibrate 10V analog programming interface for setting
current output.
•
ANLG I RB 10V: Calibrate 10V analog programming interface for
monitoring current output.
•
Factory Cal: Lets you restore the factory calibration constants.
•
Change Code: Lets you change the password code.
SCPI
To access calibration mode via remote interface, use the command:
CAL:STAT ON, "0000"
1
To check if the power supply is in calibration mode, use the command:
CAL:STAT ?
1. where the parameter "0000" may be replaced with your own 4-digit security
code.