Table C-9 System Commands
Function
SCPI Command
Description
Query
Query System Error
[:]SYSTem[
Returns the next error in the
instrument’s error queue
N/A
Recall Default
Factory Preset
[:]SYSTem[
Restore the factory preset values
N/A
Select Remote
Control Source
[:]SYSTem[
{RS232|GPIB|AVOLtage|ACURrent|AVCurrent|MCHann
el}
Select the remote control source
Yes
Set RS-232 Baud
Rate
[:]SYSTem[
ve]:BAUD {1200|2400|4800|9600|…}
Configure the RS-232 baud rate
Yes
Set RS-232 Flow
Control
[:]SYSTem[
ve]:PACE {HARDware|XON|NONE}
Select type of flow control for
RS-232
Yes
Set Multichannel
Address
(6000 W only)
[:]SYSTem[
ess
Select the multichannel unit
address
Yes
Set GPIB Address
[:]SYSTem[
DDRess
Configure GPIB address
Yes
Set GPIB Power On
Service Request
[:]SYSTem[
ONSrq {ON|OFF|0|1}
Configure GPIB PON SRQ
Yes
Select Range for
Analog
Programming
Interface
[:]SYSTem[
l {5|10}
Select analog interface voltage
levels
Yes
Set Remote Control
Operation (Serial
Interface)
[:]SYSTem[
{LOCal|REMote|RWLock}
RS-232 Only. Change remote
control mode
Yes
Query SCPI Version
[:]SYSTem[
Returns the SCPI version to
which the instrument complies.
Format is YYYY.V
N/A