Dose Information

– if the automatic dose increase for obese patients (or

patient sections) has to be more moderate than the
preset (choose obese: weak increase), resulting in
more image noise and a lower dose for those
images.

– if the automatic dose decrease for slim patients (or

patient sections) has to be stronger than the preset
(choose slim: strong decrease), resulting in more
image noise and a lower dose for those images.

– if the automatic dose decrease for slim patients (or

patient sections) has to be more moderate than the
preset (choose slim: weak decrease), resulting in
less image noise and a higher dose for those images.

On the Patient tab card you can adjust the image qual-
ity (for more information see chapter How does CARE
Dose 4D work
).

Note: Changing this adaptation strength effects all
protocols!