beautypg.com

Neck – Siemens EMOTION 16-Jun User Manual

Page 213

background image

Neck

213

Hints

• Due to its iodine content, the thyroid gland is hyper-

dense in relation to the neighboring muscles both
before and after an IV CM injection. For displays of
the parotid or thyroid gland or the floor of the
mouth, the slice thickness should be < 5 mm and the
length of the range should be adapted to match the
anatomic region.

• Target the FoV to ensure adequate coverage of the

region of interest in the upper neck & middle neck
levels as well as to include the axilla in the lower
neck level if required.

Contrast medium IV injection
Start delay

45 sec.

Flow rate

2.0 ml/sec.

Total amount

120 ml