beautypg.com

Vascular – Siemens EMOTION 16-Jun User Manual

Page 378

background image

378

Vascular

Hints

• CARE Bolus may be used to optimize the bolus tim-

ing.

• Set the ROI for monitoring scan in the aortic arch

with triggering threshold of 120 HU, or use manual
triggering.

• High quality 2D & 3D postprocessing can be achieved

using a thin slice thickness and 50% overlapping
increments.

• Thick MPRs and ThinMIPs can be created very quickly

on the 3D Task Card by just clicking on the appropri-
ate icons. The thickness of these reconstructed
images can be defined by clicking on the icons with
the right mouse to open the entry field.

Contrast medium IV injection
Start delay

15 sec.

Flow rate

4 ml/sec.

Total amount

90 ml