Thorax – Siemens EMOTION 16-Jun User Manual
Page 245
Thorax
245
Hints
• For lung cancer evaluation, this protocol can be com-
bined with protocol “Neck Routine”.
Emotion 6
ThorFast
2
nd
recon.
kV
130
Effective mAs/
Quality ref. mAs
70
Rotation Time
0.8/0.6 sec.
Acquisition
6 x 3.0 mm
Slice collimation
3.0 mm
Slice width
6.0 mm
6.0 mm
Feed/Rotation
27.0 mm
Pitch Factor
1.50
Increment
6.0 mm
6.0 mm
Kernel
B41s
B70s
CTDI
Vol
7.14 mGy
Effective dose
Male: 4.12 mSv
Female: 5.25 mSv
Contrast medium IV injection
Start delay
25 – 30 sec.
Flow rate
2.5 ml/sec.
Total amount
80 ml