Abdomen – Siemens EMOTION 16-Jun User Manual
Page 292
292
Abdomen
CARE Dose 4D is off as default because for CT Colonog-
raphy protocols the lowest mAs values are used.
For further information on the scan protocols and how
to use syngo Colonography, please refer to the Appli-
cation Guide "Clinical Applications".
Emotion 6
Colo_supine
Colo_prone
kV
130
130
Effective mAs/
Quality ref. mAs
50
30
Rotation Time
0.8/0.6 sec.
0.8/0.6 sec.
Acquisition
6 x 1.0 mm
6 x 1.0 mm
Slice collimation
1.0 mm
1.0 mm
Slice width
1.25 mm
1.25 mm
Feed/Rotation
10.8/9.0 mm
10.8/9.0 mm
Pitch Factor
1.80/1.50
1.80/1.50
Increment
0.8 mm
0.8 mm
Kernel
B41s
B20s
CTDI
Vol
6.35 mGy
3.81 mGy
Effective dose
Male:
Female:
4.61/4.53 mSv
7.01/6.96 mSv
2.77/2.72 mSv
4.20/4.18 mSv
Contrast medium IV injection
Start delay
50 – 60 sec.
Flow rate
3.0 ml/sec.
Total amount
100 ml