Abdomen

CARE Dose 4D is off as default because for CT Colonog-
raphy protocols the lowest mAs values are used.

For further information on the scan protocols and how
to use syngo Colonography, please refer to the Appli-
cation Guide "Clinical Applications".

Emotion 6

Colo_supine

Colo_prone

kV

130

130

Effective mAs/

Quality ref. mAs

50

30

Rotation Time

0.8/0.6 sec.

0.8/0.6 sec.

Acquisition

6 x 1.0 mm

6 x 1.0 mm

Slice collimation

1.0 mm

1.0 mm

Slice width

1.25 mm

1.25 mm

Feed/Rotation

10.8/9.0 mm

10.8/9.0 mm

Pitch Factor

1.80/1.50

1.80/1.50

Increment

0.8 mm

0.8 mm

Kernel

B41s

B20s

CTDI

Vol

6.35 mGy

3.81 mGy

Effective dose

Male:

Female:

4.61/4.53 mSv

7.01/6.96 mSv

2.77/2.72 mSv

4.20/4.18 mSv

Contrast medium IV injection

Start delay

50 – 60 sec.

Flow rate

3.0 ml/sec.

Total amount

100 ml