Spine
309
Emotion 16
L3-4
L4-5
L5-S1
kV
130
130
130
Effective mAs/
Quality ref.
mAs
280
280
320
Rotation time
1.5 sec.
1.5 sec.
1.5 sec.
Acquisition
12 x
0.6 mm
12 x
0.6 mm
12 x
0.6 mm
Slice
collimation
0.6 mm
0.6 mm
0.6 mm
Slice width
2.4 mm
2.4 mm
2.4 mm
Feed/Scan
7.2 mm
7.2 mm
7.5 mm
Kernel
B31s
B31s
B31s
CTDI
Vol
38.64
mGy
38.49
mGy
42.66
mGy
Effective dose
Male:
Female:
1.75 mSv
2.45 mSv
2.15 mSv
3.78 mSv
2.43 mSv
5.56 mSv