beautypg.com

Pelvis – Siemens EMOTION 16-Jun User Manual

Page 323

background image

Pelvis

323

* If the examination performed requires a full urinary
bladder, wait at least 3 minutes following IV adminis-
tration of the contrast medium.

Emotion 6

Pelvis

2

nd

reconstr.

kV

130

Effective mAs/
Quality ref. mAs

120

Rotation Time

1.0 sec.

Acquisition

6 x 2.0 mm

Slice collimation

2.0 mm

Slice width

5.0 mm

5.0 mm

Feed/Rotation

18.0 mm

Pitch Factor

1.50

Increment

5.0 mm

5.0 mm

Kernel

B41s

B41s

CTDI

Vol

13.08 mGy

Effective Dose

Male: 5.13 mSv
Female: 8.76 mSv

Contrast medium IV injection
Start delay

50 sec.*

Flow rate

2.0 – 3.0 ml/sec.

Total amount

100 – 120 ml