Pelvis – Siemens EMOTION 16-Jun User Manual
Page 323
Pelvis
323
* If the examination performed requires a full urinary
bladder, wait at least 3 minutes following IV adminis-
tration of the contrast medium.
Emotion 6
Pelvis
2
nd
reconstr.
kV
130
Effective mAs/
Quality ref. mAs
120
Rotation Time
1.0 sec.
Acquisition
6 x 2.0 mm
Slice collimation
2.0 mm
Slice width
5.0 mm
5.0 mm
Feed/Rotation
18.0 mm
Pitch Factor
1.50
Increment
5.0 mm
5.0 mm
Kernel
B41s
B41s
CTDI
Vol
13.08 mGy
Effective Dose
Male: 5.13 mSv
Female: 8.76 mSv
Contrast medium IV injection
Start delay
50 sec.*
Flow rate
2.0 – 3.0 ml/sec.
Total amount
100 – 120 ml