beautypg.com

Thorangiovol, Thorangiovol 384, Vascular – Siemens EMOTION 16-Jun User Manual

Page 384

background image

384

Vascular

ThorAngioVol

Indications:

Spiral mode for thorax CT Angio studies, e.g. visualiza-
tion of tumors, metastases, lymphoma, lymph nodes,
vascular anomalies etc.

Two recon jobs are predefined for reconstruction: the
first for axial, the second for double-oblique studies in
3D images display view.

The oblique view images will be reconstructed as MIP
images.

For SOMATOM Emotion 16-slice configuration:

A range of 30 cm will be covered in 13.70 sec.

For SOMATOM Emotion 6-slice configuration:

A range of 30 cm will be covered in 28.27 sec.

Emotion 16

ThorAngio

2

nd

reconstr.

kV

110

Effective mAs/
Quality ref. mAs

70

Rotation time

0.6 sec.

Acquisition

16 x 0.6 mm

Slice collimation

0.6 mm

Slice width

5.0 mm

3.0 mm

Feed/Rotation

14.4 mm

Pitch Factor

1.50

Increment

5.0 mm

3.0 mm

Kernel

B31s

B30s

CTDI

Vol

5.60 mGy

Effective dose

Male: 2.92 mSv
Female: 3.78 mSv