Contrast Medium

133

Additional Important Information

• If Coupled Mode is selected the CT checks if there is

an injector available.
The scan mode cannot be loaded if a connection can-
not be established or if the injector does not accept
coupling (The injector will not accept coupling while
injecting).
A message appears: Injector is not connected.

• Scanning interrupted

If the injector does not accept the START from the CT
the scan mode is cancelled.
If the scanner is suspended by the user or if technical
problems occur, the injector will be stopped too.

• Injector stopped

If the injector is stopped by the user the scan will be
stopped too.
If the injector is interrupted, by pressing the Hold
button the scan will be continued.