Testbolus protocol, Testbolus, Testbolus 460 – Siemens EMOTION 16-Jun User Manual
Page 460: Specials testbolus protocol
460
Specials
TestBolus Protocol
TestBolus
This mode can be used to test the start delay of an opti-
mal enhancement after the contrast medium injec-
tion.
Emotion 16
TestBolus
kV
110
Effective mAs/
Quality ref. mAs
40
Rotation time
0.6 sec.
Acquisition
2 x 5.0 mm
Slice collimation
5.0 mm
Slice width
10.0 mm
Feed/Scan
0.0 mm
Kernel
B31s
CTDI
vol
2.60 mGy
Emotion 6
TestBolus
kV
110
Effective mAs/
Quality ref. mAs
40
Rotation time
0.6 sec.
Acquisition
2 x 5.0 mm
Slice collimation
5.0 mm
Slice width
10.0 mm
Feed/Scan
0.0 mm
Kernel
B31s
CTDI
vol
2.48 mGy