Vascular – Siemens EMOTION 16-Jun User Manual
Page 367
Vascular
367
Emotion 16
HeadAngio
2
nd
reconstr.
kV
110
Effective mAs/
Quality ref. mAs
70
Rotation time
0.6 sec.
Acquisition
16 x 0.6 mm
Slice collimation
0.6 mm
Slice width
3.0 mm
1.0 mm
Feed/Rotation
7.7 mm
Pitch Factor
0.80
Increment
3.0 mm
0.7 mm
Kernel
H31s
H20s
CTDI
Vol
12.46 mGy
Effective dose
Male: 0.43 mSv
Female: 0.48 mSv
Emotion 6
HeadAngio
2
nd
reconstr.
kV
110
Effective mAs/
Quality ref. mAs
70
Rotation time
0.8/0.6 sec.
Acquisition
6 x 1.0 mm
Slice collimation
1.0 mm
Slice width
3.0 mm
1.25 mm
Feed/Rotation
5.1 mm
Pitch Factor
0.85
Increment
3.0 mm
0.8 mm
Kernel
H31s
H20s
CTDI
Vol
12.88 mGy
Effective dose
Male: 0.37 mSv
Female: 0.42 mSv