Spinevol, Spinevol 305, Spine – Siemens EMOTION 16-Jun User Manual
Page 305
Spine
305
SpineVol
Indications:
Spiral mode for thoracic and lumbar spine studies, e.g.
prolapse, degenerative changes, trauma, tumors etc.
Three recon jobs are predefined for reconstruction: the
first for soft tissue axial, the second for soft tissue sag-
ittal and third for coronal bone studies in 3D images
display view.
For SOMATOM Emotion 16-slice configuration:
A range of 15 cm will be covered in 26.04 sec.
For SOMATOM Emotion 6-slice configuration:
A range of 10 cm will be covered in 32.30 sec.
Emotion 16
Spine
2
nd
recon.
3
rd
recon.
kV
130
Effective mAs/
Quality ref. mAs
190
Rotation Time
1.0 sec.
Acquisition
16 x 0.6 mm
Slice collimation
0.6 mm
Slice width
3.0 mm
2.0 mm
2.0 mm
Feed/Rotation
6.2 mm
Pitch Factor
0.65
Increment
3.0 mm
2.0 mm
2.0 mm
Kernel
B31s
B20s
B20s
CTDI
Vol
23.75 mGy
Effective dose
Male: 4.42 mSv
Female: 4.70 mSv