beautypg.com

Spinevol, Spinevol 305, Spine – Siemens EMOTION 16-Jun User Manual

Page 305

background image

Spine

305

SpineVol

Indications:

Spiral mode for thoracic and lumbar spine studies, e.g.
prolapse, degenerative changes, trauma, tumors etc.

Three recon jobs are predefined for reconstruction: the
first for soft tissue axial, the second for soft tissue sag-
ittal and third for coronal bone studies in 3D images
display view.

For SOMATOM Emotion 16-slice configuration:

A range of 15 cm will be covered in 26.04 sec.

For SOMATOM Emotion 6-slice configuration:

A range of 10 cm will be covered in 32.30 sec.

Emotion 16

Spine

2

nd

recon.

3

rd

recon.

kV

130

Effective mAs/
Quality ref. mAs

190

Rotation Time

1.0 sec.

Acquisition

16 x 0.6 mm

Slice collimation

0.6 mm

Slice width

3.0 mm

2.0 mm

2.0 mm

Feed/Rotation

6.2 mm

Pitch Factor

0.65

Increment

3.0 mm

2.0 mm

2.0 mm

Kernel

B31s

B20s

B20s

CTDI

Vol

23.75 mGy

Effective dose

Male: 4.42 mSv
Female: 4.70 mSv