beautypg.com

Shoulder – Siemens EMOTION 16-Jun User Manual

Page 225

background image

Shoulder

225

Hints

• Use raw data to review a target region if necessary.
• For image reconstruction of soft tissue use kernel

B31s and a slice width of 5.0 mm.

• Coronal and sagittal 2D planar reconstructions are

important for evaluation of the joint space & bursa
sacs in CT arthograms.

• 3D renderings are helpful for complex fractures &

dislocations.

Emotion 6

Shoulder

2

nd

recon.

3

rd

recon.

kV

130

Effective mAs/
Quality ref. mAs

100

Rotation Time

1.5 sec.

Acquisition

6 x 1.0 mm

Slice collimation

1.0 mm

Slice width

5.0 mm

5.0 mm

5.0 mm

Feed/Rotation

6.0 mm

Pitch Factor

1.0

Increment

5.0 mm

5.0 mm

5.0 mm

Kernel

B60s

B20s

B20s

CTDI

Vol

12.70 mGy

Effective dose

Male: 2.96 mSv
Female: 3.52 mSv