Head
For further information on the scan protocols and how
to use syngo Dental CT, please refer to the Application
Guide "Clinical Applications".
Emotion 6
Dental
kV
130
Effective mAs/
Quality ref. mAs
45
Rotation Time
0.8 sec.
Acquisition
6 x 1.0 mm
Slice collimation
1.0 mm
Slice width
1.25 mm
Feed/Rotation
5.1 mm
Pitch Factor
0.85
Increment
0.6 mm
Kernel
H70s
CTDI
Vol
12.38 mGy
Effective dose
Male: 0.19 mSv