beautypg.com

Innerearhrvol, Innerearhrvol 180, Head – Siemens EMOTION 16-Jun User Manual

Page 180

background image

180

Head

InnerEarHRVol

Indications:

Spiral mode for high resolution inner ear studies,
e.g.malformations of the inner ear, inflammatory-
changes, pathologies of the mastoid process, tumor-
processes of the pyramids, posttraumatic changes,etc.

Three recon jobs are predefined for reconstruction: the
first for axial bone structure, the second and third for
the double oblique for each side in 3D images display
view.

For SOMATION Emotion 16-slice configuration:

A range of 40 mm will be covered in 24.22 sec.

For SOMATION Emotion 6-slice configuration:

A range of 40 mm will be covered in 17.69 sec.

Emotion 16

Inner
Ear

2

nd

recon.

3

d

recon.

kV

130

Effective mAs/

Quality ref. mAs

130

Rotation Time

1.0 sec.

Acquisition

4 x 0.6 mm

Slice collimation

0.6 mm

Slice width

2.0 mm

2.0 mm

2.0 mm

Feed/Rotation

1.1 mm

Pitch Factor

0.45

Increment

2.0 mm

2.0 mm

2.0 mm

Kernel

H90s

H90s

H90s

CTDI

Vol

38.35 mGy

Effective dose

Male: 0.74 mSv

Female: 0.78 mSv