Osteo, Osteo 312, Spine – Siemens EMOTION 16-Jun User Manual
Page 312
312
Spine
Osteo
This is the scan protocol for the syngo Osteo CT appli-
cation package to assist the physician with the quanti-
tative assessment of vertebral bone mineral density
(BMD) in the diagnosis and follow-up of osteopenia
and osteoporosis.
Emotion 16
Osteo
kV
80
Effective mAs/
Quality ref. mAs
220
Rotation time
1.0 sec.
Acquisition
2 x 5.0 mm
Slice collimation
5.0 mm
Slice width
10.0 mm
Feed/Scan
0.0 mm
Kernel
S80s
CTDI
Vol
5.72 mGy
Effective dose
Male: 0.48 mSv
Female: 0.71 mSv
Emotion 6
Osteo
kV
80
Effective mAs/
Quality ref. mAs
220
Rotation time
1.0 sec.
Acquisition
2 x 5.0 mm
Slice collimation
5.0 mm
Slice width
10.0 mm
Feed/Scan
0.0 mm
Kernel
S80s
CTDI
Vol
5.50 mGy
Effective dose
Male: 0.21 mSv
Female: 0.49 mSv