Children – Siemens EMOTION 16-Jun User Manual
Page 563
Children
563
* The conversion factor for a 7-year-old child, and a
scan range of 160 mm was used.
Emotion 16
Spine
2
nd
reconstr.
kV
110
Effective mAs/
Quality ref. mAs
60
Rotation time
1.0 sec.
Acquisition
16 x 0.6 mm
Slice collimation
0.6 mm
Slice width
2.0 mm
1.0 mm
Feed/Rotation
6.2 mm
Pitch Factor
0.65
Increment
2.0 mm
0.7 mm
Kernel
B41s
B60s
CTDI
Vol
4.80 mGy
Effective dose
Male: 1.28 mSv*
Female: 1.43 mSv*
Emotion 6
SpineRoutine
2
nd
reconstr.
kV
110
Effective mAs/
Quality ref. mAs
78
Rotation time
1.0 sec.
Acquisition
6 x 2.0 mm
Slice collimation
2.0 mm
Slice width
3.0 mm
2.5 mm
Feed/Rotation
10.2 mm
Pitch Factor
0.85
Increment
3.0 mm
1.5 mm
Kernel
B41s
B60s
CTDI
Vol
5.54 mGy
Effective dose
Male: 2.36 mSv*
Female: 4.86 mSv*