Headangiovol, Headangiovol 370, Vascular – Siemens EMOTION 16-Jun User Manual
Page 370
370
Vascular
HeadAngioVol
Indications:
Spiral mode for cerebral CT Angio studies, e.g. cerebral
vascu-lar abnormalities, tumors and follow-up studies
etc.
Two recon jobs are predefined for reconstruction: the
first for axial, the second for coronal studies in 3D
images display view.
The coronal view images will be reconstructed as MIP
images.
For SOMATOM Emotion 16-slice configuration:
A range of 8.0 cm will be covered in 7.45 sec.
For SOMATOM Emotion 6-slice configuration:
A range of 8.0 cm will be covered in 14.15 sec.
Emotion 16
HeadAngio
2
nd
reconstr.
kV
110
Effective mAs/
Quality ref. mAs
70
Rotation time
0.6 sec.
Acquisition
16 x 0.6 mm
Slice collimation
0.6 mm
Slice width
3.0 mm
4.0 mm
Feed/Rotation
7.7 mm
Pitch Factor
0.80
Increment
3.0 mm
4.0 mm
Kernel
H31s
H20s
CTDI
Vol
12.46 mGy
Effective dose
Male: 0.43 mSv
Female: 0.48 mSv