Children – Siemens EMOTION 16-Jun User Manual
Page 569
Children
569
* The conversion factor for a 7-year-old child, and a
scan range of 60 mm was used.
** Adjust the mAs value to the body region.
Hints
• Children with a body weight of more than 55 kg
should be examined with an adult protocol.
• To work without CARE Dose 4D use for children
< 25 kg 26 mAs
25-34 kg 40 mAs
35-54 kg 80 mAs
Emotion 6
ExtrHR
kV
110
Effective mAs/
Quality ref. mAs**
50
Rotation time
1.0 sec.
Acquisition
6 x 0.5 mm
Slice collimation
0.5 mm
Slice width
0.6 mm
Feed/Rotation
3.0 mm
Pitch Factor
1.00
Increment
0.3 mm
Kernel
U90s
CTDI
Vol
4.25 mGy
Effective dose
Male: 0.01 mSv*
Female: 0.01 mSv*