beautypg.com

Pelvisvol, Pelvisvol 322, Pelvis – Siemens EMOTION 16-Jun User Manual

Page 322

background image

322

Pelvis

PelvisVol

Indications:

Spiral mode for pelvis studies, e.g. processes of the
prostate, urinary bladder, rectum, gynecological indi-
cations etc.

Two recon jobs are predefined for reconstruction: the
first for axial, the second for coronal studies in 3D
images display view.

For SOMATOM Emotion 16-slice configuration:

A typical range of 20 cm will be covered in 8.94 sec.

For SOMATOM Emotion 6-slice configuration:

A typical range of 20 cm will be covered in 13.11 sec.

Emotion 16

Pelvis

2

nd

reconstr.

kV

130

Effective mAs/
Quality ref. mAs

120

Rotation Time

1.0 sec.

Acquisition

16 x 1.2 mm

Slice collimation

1.2 mm

Slice width

5.0 mm

5.0 mm

Feed/Rotation

28.8 mm

Pitch Factor

1.50

Increment

5.0 mm

5.0 mm

Kernel

B41s

B41s

CTDI

Vol

13.44 mGy

Effective Dose

Male: 5.32 mSv
Female: 9.14 mSv