Rt_abdomen, Rt_abdomen 478, Radiation therapy – Siemens EMOTION 16-Jun User Manual
Page 478
478
Radiation Therapy
RT_Abdomen
Indications:
Spiral mode for routine radiation therapy planning
abdominal studies.
For SOMATOM Emotion 16-slice configuration:
A range of 20 cm will be covered in 9.01 sec.
For SOMATOM Emotion 6-slice configuration:
A range of 20 cm will be covered in 17.29 sec.
Emotion 16
Abdomen
2
nd
recon.
kV 130
Effective mAs/
Quality ref. mAs
150
Rotation time
0.6 sec.
Acquisition
16 x 1.2 mm
Slice collimation
1.2 mm
Slice width
8.0 mm
8.0 mm
Feed/Rotation
15.4 mm
Pitch Factor
0.80
Increment
8.0 mm
8.0 mm
Kernel
B41s
B60s
CTDI
Vol
16.80 mGy
Effective dose
Male: 7.36 mSv
Female: 8.94 mSv