Hipvol, Hipvol 328, Pelvis – Siemens EMOTION 16-Jun User Manual

Page 328

background image

328

Pelvis

HipVol

Indications:

Spiral mode for bone studies and soft tissue studies of
the Hip, e.g. evaluation of joint cavity, masses, trauma,
dysplasia, necrosis of the head of the hip, congruence
evaluations, orthopedic indications etc.

Two recon jobs are predefined for reconstruction: the
first for axial, the second for coronal studies in 3D
images display view.

For SOMATOM Emotion 16-slice configuration:

A typical range of 10 cm will be covered in 8.94 sec.

For SOMATOM Emotion 6-slice configuration:

A typical range of 10 cm will be covered in 13.11 sec.

Two recon jobs are predefined for reconstruction: the
first for axial, the second for coronal studies in
3D images display view.

Emotion 16

Hip

2

nd

reconstr.

kV

130

Effective mAs/
Quality ref. mAs

100

Rotation time

1.0 sec.

Acquisition

16 x 0.6 mm

Slice collimation

0.6 mm

Slice width

5.0 mm

5.0 mm

Feed/Rotation

14.4 mm

Pitch Factor

1.50

Increment

5.0 mm

5.0 mm

Kernel

B70s

B60s

CTDI

Vol

12.50 mGy

Effective Dose

Male: 2.31 mSv
Female: 5.93 mSv