Hipvol, Hipvol 328, Pelvis – Siemens EMOTION 16-Jun User Manual
Page 328
328
Pelvis
HipVol
Indications:
Spiral mode for bone studies and soft tissue studies of
the Hip, e.g. evaluation of joint cavity, masses, trauma,
dysplasia, necrosis of the head of the hip, congruence
evaluations, orthopedic indications etc.
Two recon jobs are predefined for reconstruction: the
first for axial, the second for coronal studies in 3D
images display view.
For SOMATOM Emotion 16-slice configuration:
A typical range of 10 cm will be covered in 8.94 sec.
For SOMATOM Emotion 6-slice configuration:
A typical range of 10 cm will be covered in 13.11 sec.
Two recon jobs are predefined for reconstruction: the
first for axial, the second for coronal studies in
3D images display view.
Emotion 16
Hip
2
nd
reconstr.
kV
130
Effective mAs/
Quality ref. mAs
100
Rotation time
1.0 sec.
Acquisition
16 x 0.6 mm
Slice collimation
0.6 mm
Slice width
5.0 mm
5.0 mm
Feed/Rotation
14.4 mm
Pitch Factor
1.50
Increment
5.0 mm
5.0 mm
Kernel
B70s
B60s
CTDI
Vol
12.50 mGy
Effective Dose
Male: 2.31 mSv
Female: 5.93 mSv