Shoulder

ShoulderVol

Indications:

Spiral mode for bone studies and soft tissues, e.g. eval-
uation of joint cavities, masses, trauma, dislocations,
orthopedic indications etc.

Three recon jobs are predefined for reconstruction: the
first for soft tissue axial, the second for soft tissue coro-
nal and third for sagittal bone studies in 3D images dis-
play view.

For SOMATOM Emotion 16-slice configuration:

A scan range of 15 cm will be covered in 26.44 sec.

For SOMATOM Emotion 6-slice configuration:

A scan range of 15 cm will be covered in 40.50 sec.

Emotion 16

Shoulder

2

nd

recon.

3

rd

recon.

kV

130

Effective mAs/
Quality ref. mAs

100

Rotation Time

1.5 sec.

Acquisition

16 x 0.6 mm

Slice collimation

0.6 mm

Slice width

5.0 mm

5.0 mm

5.0 mm

Feed/Rotation

9.6 mm

Pitch Factor

1.0

Increment

5.0 mm

5.0 mm

5.0 mm

Kernel

B60s

B31s

B60s

CTDI

Vol

12.50 mGy

Effective dose

Male: 3.38 mSv
Female: 3.75 mSv