Spinethinslice, Spinethinslice 304, Spine – Siemens EMOTION 16-Jun User Manual
Page 304
304
Spine
SpineThinSlice
Indications:
Spiral mode for thin slice thoracic and lumbar spine
studies, e.g. prolapse, degenerative changes, trauma,
tumors etc.
For SOMATOM Emotion 6-slice configuration:
A range of 10 cm will be covered in 32.30 sec.
F
o
r
S
O
M
A
T
Emotion 6
Spine
ThinSlice
2
nd
recon.
3
rd
recon.
kV
130
Effective mAs/
Quality ref. mAs
190
Rotation Time
1.0 sec.
Acquisition
6 x 1.0 mm
Slice collimation
1.0 mm
Slice width
3.0 mm
1.25 mm
1.25 mm
Feed/Rotation
3.3 mm
Pitch Factor
0.55
Increment
3.0 mm
0.8 mm
0.8 mm
Kernel
B31s
B31s
B70s
CTDI
Vol
24.13 mGy
Effective dose
Male: 3.64 mSv
Female: 7.23 mSv