543

* The conversion factor for a 7-year-old child, and a

scan range of 170 mm was used.

Emotion 16

Neck

2

nd

reconstr.

kV

110

Effective mAs/
Quality ref. mAs

50

Rotation time

1.0 sec.

Acquisition

16 x 1.2 mm

Slice collimation

1.2 mm

Slice width

5.0 mm

1.5 mm

Feed/Rotation

28.8 mm

Pitch Factor

1.50

Increment

5.0 mm

1.0 mm

Kernel

B50s

B50s

CTDI

Vol

3.60 mGy

Effective dose

Male: 0.69 mSv*
Female: 0.71 mSv*

Emotion 6

Neck

2

nd

reconstr.

kV

110

Effective mAs/
Quality ref. mAs

50

Rotation time

0.8 sec.

Acquisition

6 x 2.0 mm

Slice collimation

2.0 mm

Slice width

5.0 mm

2.5 mm

Feed/Rotation

18.0 mm

Pitch Factor

1.50

Increment

5.0 mm

1.5 mm

Kernel

B50s

B50s

CTDI

Vol

3.55 mGy

Effective dose

Male: 2.03 mSv*