Polytrauma/ polytrauma06s, Polytrauma, Polytrauma06s 426 – Siemens EMOTION 16-Jun User Manual
Page 426: Specials
426
Specials
PolyTrauma/
PolyTrauma06s
Two combined ranges are predefined, head with neck
and thorax with abdomen.
For SOMATOM Emotion 16-slice configuration:
A scan range of 12/14 cm will be done in 13.36/9.68
sec.
Emotion 16
Head
Neck
kV 130
130
Effective mAs/
Quality ref. mAs
240
150
Rotation time
1.0 sec.
1.0 sec
Acquisition
16 x 1.2 mm
16 x 1.2 mm
Slice collimation
1.2 mm
1.2 mm
Slice width
6.0 mm
5.0 mm
Feed/Rotation
10.6 mm
19.2 mm
Pitch Factor
0.55
1.00
Increment
6.0 mm
5.0 mm
Kernel
H31s
B50s
CTDI
Vol
57.60 mGy
16.80 mGy
Effective dose
Male:
Female:
2.63 mSv
2.88 mSv
3.08 mSv
3.37 mSv