Spinethinslice, Spinethinslice 566, Children – Siemens EMOTION 16-Jun User Manual
Page 566
566
Children
SpineThinSlice
Indications:
Spiral mode for the spine when Multi Planar Reforma-
tion (MPR) are intended, for example, post-traumatic
changes, tumors, malformations, etc.
For SOMATOM Emotion 6-slice configuration:
A typical range of 15 cm covered in 47.45 sec.
* The conversion factor for a 7-year-old child, and a
scan range of 160 mm was used.
Emotion 6
SpineThinSlice
2
nd
reconstr.
kV
110
Effective mAs/
Quality ref. mAs
78
Rotation time
1.0 sec.
Acquisition
6 x 1.0 mm
Slice collimation
1.0 mm
Slice width
2.0 mm
1.25 mm
Feed/Rotation
3.3 mm
Pitch Factor
0.55
Increment
2.0 mm
0.8 mm
Kernel
B41s
B60s
CTDI
Vol
6.47 mGy
Effective dose
Male: 2.55 mSv*
Female: 3.02 mSv*