beautypg.com

Bodyangiovol, Bodyangiovol 402, Vascular – Siemens EMOTION 16-Jun User Manual

Page 402

background image

402

Vascular

BodyAngioVol

Indications:

Spiral mode for abdominal CT Angio studies.

Two recon jobs are predefined for reconstruction: the
first for axial, the second for coronal studies in 3D
images display view. The coronal view images will be
reconstructed as MIP images.

For SOMATOM Emotion 16-slice configuration:

A typical study in a range of 40 cm will be covered in
17.87 sec.

For SOMATOM Emotion 6-slice configuration:

A typical study in a range of 20 cm will be covered in
19.38 sec.

Emotion 16

BodyAngio

2

nd

reconstr.

kV

110

Effective mAs/

Quality ref. mAs

90

Rotation Time

0.6 sec.

Acquisition

16 x 0.6 mm

Slice collimation

0.6 mm

Slice width

5.0 mm

3.0 mm

Feed/Rotation

14.4 mm

Pitch Factor

1.50

Increment

5.0 mm

3.0 mm

Kernel

B31s

B30s

CTDI

Vol

7.20 mGy

Effective dose

Male: 5.04 mSv

Female: 6.36 mSv