Children – Siemens EMOTION 16-Jun User Manual
Page 579
Children
579
* The conversion factor for a 7-year-old child, and a
scan range of 170 mm was used.
Emotion 16
CarotidAngio
2
nd
reconstr.
kV
80
Effective mAs/
Quality ref. mAs
40
Rotation time
0.6 sec.
Acquisition
16 x 0.6 mm
Slice collimation
0.6 mm
Slice width
3.0 mm
1.0 mm
Feed/Rotation
14.4 mm
Pitch Factor
1.50
Increment
3.0 mm
0.7 mm
Kernel
B41s
B31s
CTDI
Vol
1.28 mGy
Effective dose
Male: 0.29 mSv*
Female: 0.28 mSv*
Emotion 6
CarotidAngio
2
nd
reconstr.
kV
80
Effective mAs/
Quality ref. mAs
60
Rotation time
0.8/0.6 sec.
Acquisition
6 x 1.0 mm
Slice collimation
1.0 mm
Slice width
3.0 mm
1.25 mm
Feed/Rotation
9.0 mm
Pitch Factor
1.50
Increment
3.0 mm
0.8 mm
Kernel
B41s
B31s
CTDI
Vol
1.98 mGy
Effective dose
Male: 1.79 mSv*
Female: 1.83 mSv*