Neckvol, Neckvol 214, Neck – Siemens EMOTION 16-Jun User Manual
Page 214
214
Neck
NeckVol
Indications:
Spiral mode for soft tissue studies in the cervical
region, e.g. tumors, lymphoma, abscesses etc.
Three recon jobs are predefined for reconstruction: the
first for axial, the second for coronal and third for sagg-
ittal studies in 3D images display.
For SOMATOM Emotion 16-slice configuration:
A typical range of 20 cm will be covered in 28.04 sec.
For SOMATOM Emotion 16-slice configuration:
A typical range of 20 cm will be covered in 32.97 sec.
Emotion 16
Neck
2
nd
recon.
3
rd
recon.
kV
130
Effective mAs/
Quality ref. mAs
130
Rotation Time
1.0 sec.
Acquisition
16 x 0.6 mm
Slice collimation
0.6 mm
Slice width
5.0 mm
5.0 mm
5.0 mm
Feed/Rotation
7.7 mm
Pitch Factor
0.80
Increment
5.0 mm
5.0 mm
5.0 mm
Kernel
B50s
B20s
B20s
CTDI
Vol
16.25 mGy
Effective dose
Male: 3.61 mSv
Female: 3.90 mSv