beautypg.com

Neckvol, Neckvol 214, Neck – Siemens EMOTION 16-Jun User Manual

Page 214

background image

214

Neck

NeckVol

Indications:

Spiral mode for soft tissue studies in the cervical
region, e.g. tumors, lymphoma, abscesses etc.

Three recon jobs are predefined for reconstruction: the
first for axial, the second for coronal and third for sagg-
ittal studies in 3D images display.

For SOMATOM Emotion 16-slice configuration:

A typical range of 20 cm will be covered in 28.04 sec.

For SOMATOM Emotion 16-slice configuration:

A typical range of 20 cm will be covered in 32.97 sec.

Emotion 16

Neck

2

nd

recon.

3

rd

recon.

kV

130

Effective mAs/
Quality ref. mAs

130

Rotation Time

1.0 sec.

Acquisition

16 x 0.6 mm

Slice collimation

0.6 mm

Slice width

5.0 mm

5.0 mm

5.0 mm

Feed/Rotation

7.7 mm

Pitch Factor

0.80

Increment

5.0 mm

5.0 mm

5.0 mm

Kernel

B50s

B20s

B20s

CTDI

Vol

16.25 mGy

Effective dose

Male: 3.61 mSv
Female: 3.90 mSv