Scan protocol, Respseq, Respseq 500 – Siemens EMOTION 16-Jun User Manual
Page 500: Respiratory gating scan protocol
500
Respiratory Gating
Scan Protocol
RespSeq
Indications:
A sequential mode for studies with respiration trigger-
ing.
For SOMATOM Emotion 16-slice configuration:
The scan length is predefined with 30.5 cm.
For SOMATOM Emotion 6-slice configuration:
The scan length is predefined with 30.3 cm.
Emotion 16
RespSeq
kV
130
Effective mAs/
Quality ref. mAs
16
Rotation time
0.6 sec.
Acquisition
16 x 1.2 mm
Slice collimation
1.2 mm
Slice width
2.4 mm
Feed/Scan
19.2 mm
Kernel
B31s
CTDI
vol
1.79 mGy
Effective dose
Male: 0.92 mSv
Female: 1.18 mSv