beautypg.com

Scan protocol, Respseq, Respseq 500 – Siemens EMOTION 16-Jun User Manual

Page 500: Respiratory gating scan protocol

background image

500

Respiratory Gating

Scan Protocol

RespSeq

Indications:

A sequential mode for studies with respiration trigger-
ing.

For SOMATOM Emotion 16-slice configuration:

The scan length is predefined with 30.5 cm.

For SOMATOM Emotion 6-slice configuration:

The scan length is predefined with 30.3 cm.

Emotion 16

RespSeq

kV

130

Effective mAs/
Quality ref. mAs

16

Rotation time

0.6 sec.

Acquisition

16 x 1.2 mm

Slice collimation

1.2 mm

Slice width

2.4 mm

Feed/Scan

19.2 mm

Kernel

B31s

CTDI

vol

1.79 mGy

Effective dose

Male: 0.92 mSv
Female: 1.18 mSv