Routine subtask card, Specials – Siemens EMOTION 16-Jun User Manual

Page 458

background image

458

Specials

Routine Subtask card

Move table/scanrange to displayed image position.

The function Move Table/Scanrange to displayed
image position
in the Routine subtask card allows you
to move the table to the position of the displayed tomo
image in the selected segment. If you use the CARE
View Mode you can use the Head of the Feet image for
adjusting your table position as well.

Incremental/continuous table movement

You can switch between incremental and continuous
table movement while the scan is loaded, the default
step size will be changed according to the chosen slice
thickness (default is always half the current slice thick-
ness) but adapt the increment in 0.5 steps.

Biopsy Icon

In the Routine subtask card you will find a button for
switching "on" the biopsy Mode for every sequential
scan protocol. The box where you can define if you
want to scan cranio-caudal or caudo-cranial will then
change and you can decide if you want to move the
table with an incremental or continuous table move-
ment. Then the interventional Toolbar will be available
as well.

Move table top only

If you want to only move the Table top and have the
Top Support fixed you can use the function Move
table top only
in the Routine Card. Make sure that the
Top Support (table) is completely out.