Carotidangiovol, Carotidangiovol 376, Vascular – Siemens EMOTION 16-Jun User Manual

Page 376

background image

376

Vascular

CarotidAngioVol

Indications:

CT angiography of the carotid arteries, e.g. carotid
stenosis or occlusions, coarse plaques abnormalities of
the carotids and vertebral arteries, etc.

Two recon jobs are predefined for reconstruction: the
first for axial, the second for coronal studies in 3D
images display view.

The coronal view images will be reconstructed as MIP
images.

For SOMATOM Emotion 16-slice configuration:

A range of 20 cm including the aorta arch will be cov-
ered in 9.53 sec.

For SOMATOM Emotion 6-slice configuration:

A range of 12 cm including the aorta arch will be cov-
ered in 12.27 sec.

Emotion 16

CarotidAngio

2

nd

reconstr.

kV

110

Effective mAs/
Quality ref. mAs

70

Rotation time

0.6 sec.

Acquisition

16 x 0.6 mm

Slice collimation

0.6 mm

Slice width

5.0 mm

3.0 mm

Feed/Rotation

14.4 mm

Pitch Factor

1.50

Increment

5.0 mm

3.0 mm

Kernel

B31s

B20s

CTDI

Vol

5.60 mGy

Effective dose

Male: 1.28 mSv

Female: 1.41 mSv