Abdomen – Siemens EMOTION 16-Jun User Manual
Page 285
Abdomen
285
A range of 20 cm including liver, pancreas and kidney:
arterial phase acquired in 12.96 sec.
* CARE Bolus may be used to optimize the bolus tim-
ing. Set the ROI for monitoring scan in the abdominal
aorta at the beginning of the scan range with trig-
gering threshold of 120 HU, or use manual trigger-
ing.
Emotion 6
Venous Phase
kV
130
Effective mAs/
Quality ref. mAs
120
Rotation Time
0.8/0.6 sec.
Acquisition
6 x 2.0 mm
Slice collimation
2.0 mm
Slice width
5.0 mm
Feed/Rotation
10.2 mm
Pitch Factor
0.85
Increment
5.0 mm
Kernel
B41s
CTDI
Vol
13.08 mGy
Effective dose
Male: 5.06 mSv
Female: 6.30 mSv
Contrast medium IV injection
Start delay
20 – 25 sec.* (arterial phase)
50 – 75 sec.* (venous phase)
Flow rate
4 – 5 ml/sec.
Total amount
100 –120 ml