Head – Siemens EMOTION 16-Jun User Manual
Page 175
Head
175
Emotion 16
BaseSeq
CerebrumSeq
kV
130
130
Effective mAs/
Quality ref. mAs
270
270
Rotation Time
1.5 sec.
1.5 sec.
Acquisition
2 x 5.0 mm
2 x 8.0 mm
Slice collimation
5.0 mm
8.0 mm
Slice width
5.0 mm
8.0 mm
Feed/Scan
10 mm
16 mm
Kernel
H31s
H31s
CTDI
Vol
58.32 mGy
57.24 mGy
Effective dose
Male:
1.08 mSv
Female:
1.20 mSv
Male:
1.96 mSv
Female:
2.17 mSv
Emotion 6
BaseSeq
CerebrumSeq
kV
130
130
Effective mAs/
Quality ref. mAs
250
270
Rotation Time
1.5 sec.
1.5 sec.
Acquisition
6 x 2.0 mm
6 x 3.0 mm
Slice collimation
2.0 mm
3.0 mm
Slice width
4.0 mm
6.0 mm
Feed/Scan
12 mm
18 mm
Kernel
H31s
H31s
CTDI
Vol
58.75 mGy
59.40 mGy
Effective dose
Male:
1.11 mSv
Female:
1.15 mSv
Male:
1.94 mSv
Female:
2.13 mSv